All Site-Log series data loggers come with an USB and an AUX ports. The USB port is for connecting directly to the computer. The AUX port is for connecting to a serial communication device like a Modem or an embedded Device Server when the logger needs to be accessed remotely.

Please make sure SiteView and USB driver are installed correctly. Refer to Install SiteView and USB driver for details on the installation.

USB Port Connection

Start Communication:

SiteView application starts running and the logger icon is showing in the USB tab page of the communication panel:

Adjust USB Settings:

Each time when you selected a USB icon under USB Port list, the settings for that port will be displayed under the properties section. You may change the settings based on the communication quality. For detailed explanation of each setting please refer to Communication Panel.


 Serial Port Connection

Start Communication:

SiteView application starts running. Click “Serial Port” tab in the communication panel,

 

Adjust Serial Port Settings:

Each time when you selected a COM port icon under Serial Port list, the settings for that port will be displayed under the properties section. You may change the settings based on the communication quality. For detailed explanation of each setting please refer to Communication Panel.

Ethernet Connection

Start Communication:

 

The Device Server was pre-configured with the correct parameters. You can use Link Local and multicast DNS (also known as Bonjour) to locate the Device Server and its IP address.

Bonjour Browser can be downloaded from our website at: www.microedgeinstruments.com/resourcedownload

Once you run Bonjour Browser and select Web Server(HTTP) you will see a window similar to the following:

Write down the IP address besides the SitePlayer Telnet.

  • Launch SiteView application by double clicking on SiteView icon on the desktop:

 

SiteView application starts running. Click “Ethernet” tab in the communication panel.

  • Click “New” button to create a new Ethernet connection:

In the following pop-up dialog, enter the IP address that you obtained from the preceding step. Give a meaningful description of the connection and click “OK” button to close the dialog.

Notice a new connection icon was added to the list.

If you use other brand Device Server you may need to adjust Port and Packet Size values accordingly in order to make a successful connection.

  • Double click on the icon to start the communication. If the communication is successful the logger's status page will be displayed.

Adjust Ethernet Settings:

Each time when you selected a Ethernet icon under IP Address list, the settings for that address will be displayed under the properties section. You may change the settings based on the communication quality. For detailed explanation of each setting please refer to Communication Panel.

Modem Connection

When you use Modem to communicate with a logger, you need one local Modem to connect to the computer where SiteView is running, and another remote Modem to connect to the logger.

The following figure illustrates the correct connection between a computer and a logger.

 

The Modem settings are under the Serial Port connection:

 

 

In order to make the connection, both local and remote Modems must be initialized correctly.

 

Initialize Remote Modem:

 

If the remote Modem has not been initialized by SiteView, the Modem needs to be initialized. The default initialization string that works with US Robotics is:

 

AT&F&D0&K0&Q0\N1%C0S0=1S37=9M0&M0-V90=0E0&W

To initialize a remote Modem:

The following table explains each command in Default Initialization String:

Command

Description

AT

Get the Modoem’s attention

&F

Return the configuration to the factory settings.

&D0

Control data terminal ready (DTR). 

&D0: Ignore DTR signal.

&K0

DTE/Modem Flow Control.

&K0: Disable flow control

&Q0

Asynhchronous/Synchronous mode select.

&Q0: Select asynchronous operation in the Direct mode.

\N0

Operation Mode Control.

\N0: Select normal speed buffering mode.

%C0

Compression Control

%C0: Disable data compression.

S0= x

Read or write to S registers.

S0 = 1:  Auto answer after 1 ring.

S37 = 9:   Attempt to connect at 9600 bps speed.

M0

Control Speaker On/Off

M0: Speaker is always off.

&M0

Communication Mode.

E0

Command Echo

E0: Inhibit the echoing of commands.

&W

Store user profile.

 

NOTE:  Different Modem may have different AT commands. The default initialization string is for US Robotics 33K Modem. Please refer to Appendix -> Modem AT Commands for detai

 

Initialize Local Modem:

 

The local Modem gets initialized before SiteView tries to dial the number for the Modem connection.

The default initialization string that works with US Robotics 32K Modem is:

 

AT&F &D0 &K0 &Q0 \N1 %C0 S37=13 E0 &W

 

The default initialization string is located under each COM port bar:

 

 

You may change the string based on the Modem you are using.

 

Start Communication:

 

To communicate with the logger via Modem: